Flow cytometry is a powerhouse technique in modern cell biology and diagnostics, allowing for the rapid analysis of multiple characteristics of individual cells within a heterogeneous population. Whether you're immunophenotyping, analyzing the cell cycle, or detecting intracellular proteins, a robust and well-executed staining protocol is the cornerstone of reliable and reproducible results. This guide will walk you through the essential steps and considerations for successful flow cytometry staining, demystifying the process for both newcomers and those looking to refine their techniques.
At its core, flow cytometry involves staining cells with fluorescently labeled antibodies or dyes that bind to specific cellular components. As these cells pass one by one through a laser beam, the fluorescent molecules are excited and emit light, which is then detected and converted into data. The beauty of this technique lies in its ability to provide quantitative data on a per-cell basis for thousands to millions of cells in mere minutes.
This article will provide a general framework applicable to various flow cytometry staining methods, highlighting key decision points and best practices to help you achieve clear, high-quality data.
The quality of your starting material directly impacts the quality of your flow cytometry data. Sloppy sample preparation can lead to artifacts, poor resolution, and erroneous conclusions.
Cell Isolation and Single-Cell Suspension: The journey begins with obtaining a single-cell suspension. The method will vary depending on your source material:
Suspension Cultures: These are often the easiest, usually requiring simple harvesting and washing.
Adherent Cells: Gentle detachment methods are crucial to preserve cell surface markers and viability. Enzymatic (e.g., trypsin, accutase) or non-enzymatic (e.g., EDTA-based buffers, cell scrapers) methods can be used, but optimization is key to avoid over-digestion or mechanical stress.
Tissues: Mechanical disaggregation (e.g., mincing, mashing through a cell strainer) often combined with enzymatic digestion (e.g., collagenase, DNase) is necessary. The goal is to liberate individual cells while maintaining their integrity.
Blood: Peripheral blood mononuclear cells (PBMCs) are commonly isolated using density gradient centrifugation (e.g., Ficoll-Paque). Red blood cell lysis may also be required for whole blood analysis.
Regardless of the source, the end product must be a suspension of individual cells, free of clumps and debris, which can clog the cytometer and skew results. Filtering the cell suspension through a nylon mesh (e.g., 40-70 µm) is a highly recommended final step.
Cell Counting and Viability Assessment: Knowing your cell concentration is essential for appropriate antibody staining and consistent data acquisition. Use a hemocytometer or an automated cell counter. Crucially, assess cell viability using a method like trypan blue exclusion or a fluorescent viability dye. Aim for high viability (ideally >90-95%), as dead cells can non-specifically bind antibodies and autofluoresce, leading to high background and false positives. If including a viability dye in your staining panel (highly recommended!), choose one that is compatible with your experimental setup (e.g., fixable vs. non-fixable dyes).
Cells, particularly immune cells like macrophages and B cells, express Fc receptors (FcRs) that can bind to the Fc portion of antibodies non-specifically. This can lead to false-positive signals.
Fc Receptor Blocking: Incubating your cells with an Fc blocking reagent (e.g., purified anti-Fc receptor antibodies, excess normal serum from the same species as your primary antibodies, or commercial Fc blocking solutions) before adding your specific antibodies is a critical step. This will saturate the FcRs, preventing your labeled antibodies from binding non-specifically.
General Protein Blocking: Some protocols may also recommend using a general protein blocker, like bovine serum albumin (BSA) or fetal bovine serum (FBS) in the staining buffer, to further reduce background from other non-specific protein interactions.
This is where your cells are "painted" with fluorescent labels. The approach differs slightly depending on whether you are targeting extracellular (surface) or intracellular (internal) molecules.
Extracellular Staining: Surface Marker Detection This is generally the more straightforward method.
Antibody Incubation: Add your fluorescently conjugated antibodies, diluted to their predetermined optimal concentration (see antibody titration below), to the prepared single-cell suspension.
Incubation Conditions: Incubate typically for 20-30 minutes at 2-8°C (on ice or in a cold room) in the dark. Low temperatures and the presence of sodium azide (if compatible with downstream applications) help prevent receptor internalization and antibody shedding. Keep samples protected from light to prevent photobleaching of fluorochromes.
Intracellular Staining: Unveiling Internal Proteins To detect proteins within the cell (e.g., cytokines, transcription factors, enzymes), you must first make the cell membrane permeable to antibodies. This usually involves two key steps:
Fixation: This step cross-links proteins, stabilizing the cell structure and preserving the antigen's location and integrity. Common fixatives include paraformaldehyde (PFA). The concentration and incubation time can vary depending on the target and cell type. Some epitopes can be sensitive to certain fixatives. Note: If performing combined surface and intracellular staining, it's often recommended to stain surface markers before fixation and permeabilization, as these processes can alter some surface epitopes.
Permeabilization: After fixation, cells are treated with a permeabilization reagent (e.g., saponin, Triton X-100, or specialized commercial buffers). This creates pores in the cell membrane, allowing antibodies to enter the cytoplasm and nucleus. The choice of permeabilization buffer depends on the location of your target antigen (cytoplasmic vs. nuclear) and the nature of the antigen itself. Some protocols use combined fixation/permeabilization reagents.
Antibody Incubation (Intracellular): Once cells are permeabilized, add your fluorescently conjugated antibodies targeted against intracellular antigens. Incubation conditions are similar to extracellular staining but are usually performed in permeabilization buffer.
Choosing the Right Fluorochromes and Antibodies: For multi-color flow cytometry, careful panel design is crucial. Consider:
Antigen Density: Match brighter fluorochromes with weakly expressed antigens and dimmer fluorochromes with highly expressed antigens to maximize resolution.
Spectral Overlap: Select fluorochromes with minimal spectral overlap to reduce the need for extensive compensation. Online spectral viewers are invaluable tools for this.
Antibody Specificity and Validation: Use high-quality, well-validated antibodies specific for your target antigen and species.
Titration of Antibodies: This is a critical optimization step often overlooked. Using too much antibody increases non-specific binding and background, while too little results in a weak signal. Titrate each antibody to determine the optimal concentration that gives the best signal-to-noise ratio. This involves staining cells with a range of antibody dilutions and selecting the concentration that provides bright positive staining with minimal background on negative cells.
After antibody incubation (both extracellular and intracellular), cells must be washed to remove unbound antibodies. Insufficient washing leads to high background fluorescence, obscuring genuine signals.
Typically, cells are washed 2-3 times with an appropriate wash buffer (e.g., PBS with BSA and sodium azide).
Centrifuge cells at a low speed (e.g., 300-500 x g) for 5-7 minutes at 4°C to pellet the cells without causing damage.
Carefully aspirate or decant the supernatant without disturbing the cell pellet.
Gently resuspend the cell pellet in fresh wash buffer. Avoid vigorous vortexing, which can damage cells.
Controls are non-negotiable in flow cytometry. They are essential for setting up the instrument, validating your staining, and accurately interpreting your results.
Unstained Controls: Cells processed through the entire staining procedure but without any antibodies. These help determine the level of autofluorescence (natural fluorescence of cells) and set baseline fluorescence levels (negative gates).
Isotype Controls (for antibody staining): An antibody of the same immunoglobulin class (isotype), conjugation, and concentration as your primary antibody, but directed against an antigen not present on your cells. Isotype controls help estimate non-specific binding of the antibody itself (rather than Fc receptor binding, which should be addressed by Fc blocking). Their use can be debated, and in some cases, Fluorescence Minus One (FMO) controls are preferred.
Fluorescence Minus One (FMO) Controls (for multi-color panels): In a multi-color experiment, an FMO control includes all antibodies in your panel except for one. For example, if you have a 5-color panel, you will have 5 different FMO controls. FMO controls are crucial for accurately setting gates, especially for identifying positive versus negative populations when signals are dim or when there's significant spillover from other channels.
Compensation Controls (Single-Stain Controls): When using multiple fluorochromes, the emission spectrum of one fluorochrome can spill into the detection channel of another. Compensation is a mathematical correction for this spillover. To set compensation correctly, you need single-stain controls: cells (or compensation beads) stained with only one fluorochrome from your panel for each color being used. These must be as bright as, or brighter than, the experimental samples.
Viability Dye Controls: If using a viability dye, you'll need a sample of dead cells (e.g., heat-killed) to correctly set the gate for dead cell exclusion.
Biological Controls: Positive and negative biological controls (e.g., cells known to express or not express the target antigen, or stimulated vs. unstimulated cells) can further validate your assay.
Once your cells are stained and washed, they are ready for analysis on the flow cytometer.
Instrument Setup: Ensure the cytometer is properly calibrated and quality-controlled using standardized beads. Set initial voltages and compensation using your control samples (unstained, single-stains).
Gating Strategy: Develop a logical gating strategy to isolate your population(s) of interest. Start by gating on viable, single cells, then progressively gate on specific populations based on marker expression. Use FMO controls to guide your gate placement.
Sample Resuspension: Resuspend your final cell pellet in an appropriate sheath-like buffer (e.g., PBS or a commercial FACS buffer) at the recommended concentration for your instrument (typically 0.5×106 to 2×106 cells/mL). Ensure cells are well-resuspended and filtered immediately before acquisition to prevent clogging.
Acquisition: Acquire a sufficient number of events (cells) to ensure statistical significance, especially when analyzing rare populations.
Flow cytometry protocols often require optimization for specific cell types, antibodies, and experimental questions.
Buffer Selection: Ensure buffers are Ca++/Mg++-free if cell aggregation is an issue. The addition of EDTA can also help. Maintain appropriate pH.
Incubation Times and Temperatures: These may need adjustment. While 4°C is standard for surface staining to prevent internalization, some intracellular targets may require different conditions.
Troubleshooting: Be prepared to troubleshoot common issues like weak signals, high background, cell clumping, or poor population resolution. Many resources (including manufacturer's guidelines and online forums) provide excellent troubleshooting guides.
Consistency: Once optimized, maintain consistency in all steps of your protocol for reproducible results.
Successful flow cytometry hinges on meticulous attention to detail at every stage, from sample preparation to data acquisition. While the variety of staining protocols can seem daunting, understanding the fundamental principles behind each step empowers you to adapt and optimize procedures for your specific needs. By implementing proper controls, carefully titrating antibodies, and consistently following your optimized protocol, you can unlock the rich, multi-parameter data that makes flow cytometry an indispensable tool in research and clinical settings.
Q1: What is flow cytometry and what is it used for?
A: Flow cytometry is a sophisticated technology used to rapidly analyze multiple physical and chemical characteristics of individual cells or particles as they flow in a fluid stream through a beam of light (usually a laser). It's widely used in research and clinical diagnostics for applications such as:
Identifying and quantifying different cell types in a mixed population (immunophenotyping).
Analyzing the cell cycle (DNA content).
Detecting the expression of proteins on the cell surface or inside the cell.
Assessing cell viability and apoptosis (programmed cell death).
Sorting (isolating) specific cell populations for further study.
Q2: What is the basic principle of flow cytometry?
A: The basic principle involves three core ideas:
Single-Cell Passage: Cells, often labeled with fluorescent markers, are hydrodynamically focused to pass one by one through one or more focused laser beams.
Light Interaction: As each cell passes through the laser, it scatters the light, and any fluorescent molecules on or in the cell will emit light (fluorescence) at specific wavelengths.
Detection and Analysis: Optical detectors collect the scattered and emitted light. This light is converted into electronic signals, which are then processed by a computer to generate quantitative data about the characteristics of each individual cell.
Q3: What are the 3 main components of a flow cytometer?
A: A flow cytometer generally consists of three main integrated systems:
Fluidics System: This system transports the sample, forcing cells to pass in a single file through the laser interrogation point.
Optics System: This includes lasers to illuminate the cells, and a series of lenses and filters to collect and direct the scattered and fluorescent light to the appropriate detectors (e.g., photomultiplier tubes or photodiodes).
Electronics and Computer System: This system converts the light signals detected into digital data, performs signal processing (including compensation), and allows for data storage, analysis, and visualization.
Q4: How is flow cytometry used in immunology?
A: Flow cytometry is an indispensable tool in immunology for several key applications:
Immunophenotyping: It's extensively used to identify and quantify various immune cell populations (e.g., T cells, B cells, NK cells, monocytes, dendritic cells and their subsets) based on the expression of specific cell surface markers (CD markers).
Cytokine Detection: Intracellular staining allows for the measurement of cytokine production within specific immune cells, providing insights into their functional state.
Immune Function Assays: It can assess immune cell functions like proliferation, apoptosis, phagocytosis, and cytotoxic activity.
Disease Diagnosis and Monitoring: Flow cytometry plays a crucial role in diagnosing and monitoring various immunological disorders, including leukemias, lymphomas, and immunodeficiencies, as well as tracking immune responses to therapies and vaccines.