Splotchy or uneven bands on western blot

Nothing is more frustrating in the lab than spending two days on a Western Blot, only to pull out a splotchy, uneven, or patchy membrane. This common problem can make bands unquantifiable and experiments unpublishable. But before you throw your gel box out the window, understand this: the problem is almost always solvable.



Uneven bands are not a single issue; they are a symptom. The cause could lie in your sample prep, your gel, your transfer, or your antibody steps. The key is to systematically identify the source. This article summarizes expert advice and troubleshooting guides to help you diagnose and fix the cause of your splotchy Western Blot.


Cause 1: Problems During Sample Prep and Loading Western Blot

The first errors often happen before the gel is even run. If your bands are uneven in intensity or "drag" down the lane, look here first.

  • Insoluble Debris: When you boil your samples in Laemmli buffer, proteins can denature and aggregate. If you load this insoluble matter, it can clog the wells, causing streaking or uneven entry into the stacking gel.

    • The Fix: Always centrifuge your samples at high speed (10,000+ x g) for 5-10 minutes after boiling and before loading. Only load the supernatant.

  • Uneven Loading: This seems simple, but it's a common culprit. Inaccurate protein quantification (like a BCA or Bradford assay) or simple pipetting errors will lead to bands of varying intensity.

    • The Fix: Be meticulous with your quantification and ensure all samples are in the same buffer. When loading, rinse each well 1-2 times with running buffer using a gel-loading pipette tip to clear any residual acrylamide or buffer salts.

  • High Salt Concentration: If your sample buffer has a drastically different ionic strength from the running buffer (e.t., from an IP elution), it can distort the electric field, causing bands to "smile" or "frown."

    • The Fix: Ensure your sample buffer composition is compatible with your running buffer. If necessary, use a desalting spin column to normalize your samples.


Cause 2: Issues with Gel Electrophoresis (SDS-PAGE)

If your bands look "wavy," "smiling" (edges run slower), or "frowning" (center runs slower), the problem is likely with the gel (SDS-PAGE) itself.

  • Uneven Gel Polymerization: This is a major issue with hand-cast gels. If the resolving gel polymerizes unevenly, it will create "channels" of different densities, causing proteins to migrate at different speeds.

    • The Fix: Ensure your APS and TEMED are fresh. Mix the gel solution thoroughly but gently (avoid bubbles) before pouring, and let it polymerize completely on a perfectly level surface.

  • Overheating: Gels generate heat during the run. If the tank overheats (common when running at high voltage), the center of the gel becomes warmer than the edges. This lowers the buffer's viscosity, causing the center lanes to run faster ("frowning").

    • The Fix: Run the gel at a lower constant voltage (e.g., 80-100V) for a longer time. Run the tank in a cold room (4°C) or place an ice pack in the buffer tank.

  • Expired Gels: If you use pre-cast gels, check the expiration date. An old gel can have a degraded stacking/resolving gel interface, leading to poor band stacking and smeared results.


Cause 3: The Transfer Trap: Air Bubbles and Poor Contact

This is arguably the most common cause of splotchy, patchy blots, especially "no-signal" white spots or areas of weak transfer. The proteins are separating fine but are not moving to the membrane correctly.

  • Air Bubbles: Any air bubble, no matter how small, trapped between the gel and the membrane will completely block the transfer of proteins. This results in perfect, round, white splotches on your blot where no protein is present.

    • The Fix: Be meticulous. Assemble your transfer "sandwich" (filter paper, gel, membrane, filter paper) in a tray filled with transfer buffer. This disples air. Use a "blot roller" or a clean glass pipette to gently roll out any visible bubbles from each layer.

  • Poor Membrane Activation (PVDF): PVDF membranes are hydrophobic and must be activated, or they will not bind protein.

    • The Fix: Submerge the PVDF membrane in 100% methanol for 15-30 seconds until it turns slightly opaque. Immediately transfer it to your transfer buffer and do not let it dry out.

  • Uneven Pressure/Poor Contact: If the sponges in your transfer cassette are old, compressed, or the sandwich is packed unevenly, the pressure across the gel will be inconsistent. This leads to patchy transfer, where some areas transfer well and others are faint.

    • The Fix: Use fresh, correctly-sized filter paper and sponges. Ensure the cassette closes firmly but not to the point of warping the gel.


Cause 4: Antibody Incubation and Washing Errors

If your bands are sharp but the background is splotchy, or you have tiny black dots ("freckles"), the problem is with your antibodies or washing steps.

  • Antibody Aggregation: Antibodies, especially HRP-conjugated secondaries, can form small aggregates over time. These aggregates stick randomly to the membrane, creating small, dark spots.

    • The Fix: Before diluting your primary and secondary antibodies, spin the stock vial in a microcentrifuge for 1-2 minutes to pellet any aggregates. Pipette the antibody from the top.

  • Insufficient Blocking/Washing: If your membrane is not blocked properly or washed sufficiently, the antibody will bind non-specifically, creating a high, uneven, or splotchy background.

    • The Fix: Ensure your blocking agent (e.g., 5% non-fat milk or BSA) is filtered and fresh. Most importantly, ensure your membrane is fully submerged and agitating freely during all incubation and wash steps. If the membrane sticks to the side of the container, it will develop unevenly. Use a larger volume of wash buffer (TBST) and a reliable orbital shaker.


Conclusion: A Systematic Approach

A splotchy Western Blot is a puzzle. By thinking systematically through the protocol—from Sample to Gel to Transfer to Antibodies—you can identify the step that failed. More often than not, the solution is simple: slow down, centrifuge your samples, roll out your bubbles, and make sure your membrane is always swimming freely.